User's Manual

Final documentation:
Document the findings of the survey on the floor plan. Provide one copy with the TIP book for future
review should this be necessary.
e. Completing the site survey:
Site Survey Example 1: Concerns and Logical Placement
The following site survey illustrates concerns and logical placement. This would be the first step in
designating test areas. The drawing should also show the location, or potential location, of any microwave
ovens. Also noted must be any other wireless devices in the 2.4 GHz band.
Emergency Room shown below has ordered a quantity of ten 90309's equipped with option W (wireless).
The coverage area defined by the hospital staff is any of the following areas:
Emergency rooms
Trauma rooms
One observation area
X-ray room does not require coverage
All hallways in the emergency ward
The floor plan must show desired coverage of each room to be monitored. Also listed must be hallways,
type or walls, windows, doors and the general attributes of the area where the wireless is to be installed.
Part Number 062-0983-01 Rev C page 3 - 22