User's Manual

Page 35 of 38
Use of this equipment adjacent to or stacked with other equipment (see below table) should be
avoided because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
Test Method
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
per IEC 61000-4-2
± 8 kV contact; ±
2 kV, ± 4 kV, ± 8
kV, ± 15 kV air*
,
**
±8 kV contact;
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Radiated RF EM
field immunity
per IEC 61000-4-3
10 V/m, 80 MHz
2.7 GHz, 80 % AM
at 1 kHz*
10 V/m, 80
MHz 2.7 GHz,
80 % AM at 1
kHz
NA
10 V/m, 26 MHz
2.5 GHz**
10 V/m, 26
MHz 2.5 GHz
NA
Power frequency
(50/60 Hz)
magnetic field
immunity per IEC
61000-4-8
30 A/m*
30 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital
environment.
RF wireless
communications
equipment fields
Immunity per IEC
61000-4-3
Frequencies and
levels as specified
at IEC 60601-1-2,
Table 9*
Max. 28 V/m
NA
* The pump was tested according to the EMC requirements of IEC 60601-1-2 (fourth edition)
** The pump was tested according to the EMC requirements of ISO 11608-1 (third edition)