User's Manual

Page 34 of 38
The following list provides guidelines about environmental conditions and situations to be avoided when
working with or storing the Wearable Infusion Pump:
Avoid locations where there is inadequate ventilation.
Avoid locations where sudden impact or vibration may occur.
Avoid damp locations or locations where moisture level may increase considerably.
Avoid locations with large temperature fluctuations.
Avoid locations near an electrical heating apparatus.
Avoid locations exposed to chemicals or explosive gases.
6.3 Pump Ranges
Parameter
Range
Increments
Volume To Be Infused (mL)
0.1-3
0.1
Rate (mL/h)
0.01-60
0.01
Time
00:01-05:59
1 min / 1 hour
6.4 Electromagnetic Compatibility Statement
The following sections provide information about testing of and recommendations for electromagnetic
compatibility statement.
6.4.1 Electromagnetic Emission
The Wearable Infusion Pump is intended for use in the electromagnetic environment specified below. The
customer or the user of the pump should ensure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic Environment Guidance
RF emission
CISPR 11 class B
The pump is suitable for use in home
healthcare environment.
6.4.2 Electromagnetic Immunity
The Wearable Infusion Pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the pump should ensure that it is used in the following environment.