User's Manual
Page 33 of 38
Patient population
Adult and pediatric
Delivery Mode
Continuous
Delivery route
Subcutaneous
System delivery accuracy
(Nominal)
±5%
Sensors
Air sensor, cartridge sensor, temperature sensor
Occlusion alarm threshold
4 bar
Maximum time for
activation of the occlusion
alarm
1 mL/h: 02:27 [min:sec], 1.5 mL/h: 02:05 [min:sec]
History log
Yes
ELECTRICAL SAFETY
Electrical Safety
Compliant with IEC 60601-1 - Edition 3.1
Electromagnetic
compatibility
Compliant with IEC 60601-1-2 – Edition 4.0
Recovery time for
defibrillation-proof applied
parts
< 1 sec
MECHANICAL AND POWER SPECIFICATIONS
Pump Size
23 x 95 x 46 mm HxWxD (0.91 x 3.74 x 1.81 inches)
Pump Weight
70 grams
Ingress Protection
IP65
Power Source
Non-rechargeable Li- Manganese battery 3.0 V/1000 mA
Battery consumption
6 hours of work at 0.5 mL/h
OPERATING ENVIRONMENT
Temperature
+5°C to 40°C (41°F to 104°F)
Relative Humidity
15% to 95%
Atmospheric pressure
50 kPa to 106 kPa (500 hPa to 1060 hPa)
TRANSPORT ENVIRONMENT
Temperature
-40°C to + 70°C (-40°F to 158°F)
Relative Humidity
20% to 85%
Atmospheric pressure
69.6 kPa to 106 kPa (696 hPa to 1060 hPa)
STORAGE ENVIRONMENT
Temperature
2°C to 27°C (36°F to 81°F)
Relative Humidity
20% to 85%
Atmospheric pressure
69.6 kPa to 106 kPa (696 hPa to 1060 hPa)