5 MAINTENANCE AND STORAGE
This device is for single use only! Do not try to reuse, clean or disinfect it.
5.1 Battery Classification
The UL 1642 Standard for Lithium batteries classifies the Lithium-Manganese battery used in the Wearable
Infusion Pump as primary battery (non-rechargeable).
5.2 Transport and Storage
The pump should always be transported in its single pack. During handling and transport, protect the pump and
the single pack from water, excessive humidity, and heat sources.
To safeguard the pump against prolonged exposure to dust and moisture, the pump must be stored in a clean
and dry environment.
Specific recommendations for transport and long term storage conditions are listed in the following table.
Condition
Parameters
Transport (transient up to 24
hours)
Long term storage
Temperature
-40ºC (-40ºF) to +70ºC (+158ºF)
2ºC (36ºF) to 27ºC (81ºF)
Relative humidity
20% RH to 85% RH
20% RH to 85% RH
Atmospheric pressure
69.6 kPa to 106 kPa (696 hPa to
1060 hPa)
69.6 kPa to 106 kPa (696 hPa to
1060 hPa)