User's Manual

48 US-ENGLISH
17.2. HOLTER FUNCTION
The Holter records 14 tachyarrhythmia episodes as well as the
therapy history.
STORED EPISODES
PARADYM RF VR 9250 stores 14 episodes (VF, VT, Slow VT,
SVT/ST, nonsustained).
For each episode four levels of details are presented:
Tachogram
Event log for the entire episode:
"Discrimination" analysis for each majority,
Delivered therapies,
Markers: Ventricular markers, sensed, paced and in relative
refractory periods,
EGM: onset and detection of the arrhythmia, on two therapies,
and the return to slow rhythm by recording electrogram.
Therapy history: For each arrhythmia detection, each therapy
delivered (either automatically or during an electrophysiological
study) and at the end of each arrhythmia, PARADYM RF VR 9250
records the type of majority rhythm, the number of ATP sequences
delivered, the energy and the number of shocks delivered..