User's Manual
US-ENGLISH – 33
The connector head has three ports: ventricular bipolar pace/sense
and two ports for RV and SVC defibrillation coils. The pace/sense
port is compatible with the IS-1 standard and both defibrillation ports
are compatible with the DF-1 standard. Distal lead terminal
connections are secured with set-screws accessed via self-sealing
silicone plugs. All lead connections pass through the header into the
device via feedthroughs.
Programming System: The Sorin CRM programmer is used in
conjunction with specific programmer software to interrogate and
program the implanted device at implant and during patient follow-up
procedures.
Remote Monitoring: The PARADYM RF VR 9250 is also equipped
with the RF wireless technology which enables to remotely monitor
the patients who have the Sorin CRM SMARTVIEW Monitor installed
at home.
14. IMPLANT PROCEDURE
14.1. NECESSARY EQUIPMENT
Implantation of PARADYM RF VR 9250 requires the following
equipment:
Sorin CRM dedicated programmer, equipped with the
SMARTVIEW software interface and with the programming head,
pacing system analyser, as well as its sterile connecting cables,
to evaluate the pacing and sensing thresholds,
a ventricular pacing and defibrillation lead,
physiological signal monitor capable of displaying simultaneously
the surface ECG and arterial pressure,
an external defibrillator with sterile external paddles,
sterile cover for the telemetry head.