21. NON PROGRAMMABLE PARAMETERS
21. NON PROGRAMMABLE PARAMETERS
Ventricular refractory periods Values
Post ventricular sensing 95 ms (± 16 ms)
Post ventricular pacing 220 ms (± 4 ms)
Therapies Values
Waveform
(1)
Constant tilt (50% - 50%)
Stored energy for the Max. shock 42 J
Pacing amplitude during ATP
therapies
7 V (Actual value at 300 ms: 5.3 V )
(1) The device has 50% tilt in each phase thus delivers 94% of stored energy. Each phase is limited to
10 ms duration.
SORIN – INTENSIA VR 124 – U153A