User Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Conformance to standards
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Lead connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Physical characteristics
- 19. Electrical characteristics
- 20. Programmable parameters
- 21. Non programmable parameters
- 22. Limited warranty
- 23. Patents
- 24. Explanation of symbols

20. PROGRAMMABLE PARAMETERS
─ System integrity
Lead Alerts Values
Abnormal lead impedance ON-OFF
Abnormal lead low limit (Ohm) 200-250-300-350-400-450-500
Abnormal lead high limit (Ohm) 1500-1750-2000-2500-3000
Abnormal RV coil impedance ON-OFF
Abnormal SVC coil impedance ON-OFF
Abnormal Shock impedance
(1)
ON-OFF
(1) Normal impedance range [20 Ohm-200 Ohm]
Clinical status Values
V oversensing ON-OFF
Therapy information Values
Shock disabled ON-OFF
Shocks delivered OFF-All shocks-Inefficient shock-Inefficient max shock
52
SORIN – INTENSIA VR 124 – U153A