User Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Conformance to standards
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Lead connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Physical characteristics
- 19. Electrical characteristics
- 20. Programmable parameters
- 21. Non programmable parameters
- 22. Limited warranty
- 23. Patents
- 24. Explanation of symbols

20. PROGRAMMABLE PARAMETERS
20. PROGRAMMABLE PARAMETERS
measured at 37 °C under a 500 ohm load
Legend:
Value in bold: "as shipped" value
Underlined value: nominal value
20.1. ANTIBRADYCARDIA PACING
Basic parameters Values
Mode VVI-VVIR
Basic rate (bpm)
(1)
From 30 to 90 by steps of 5 ; 60 (± 4 %)
Maximum rate (bpm) From 100 to 145 by steps of 5 ; 120 (± 6 %)
Rate hysteresis (%) 0-5-10-20-35 (± 18 ms)
(1) The corresponding periods are (in ms): 2000-1714-1500-1333-1200-1091-1000-923-857-800-750-
706-667 ms.
Special features Values
Smoothing OFF-Very slow-Slow-Medium-Fast
Physical activity Very low-Low-Medium-High-Very high
Pacing/Sensing Values
Ventricular sensitivity (mV)
(1)
From 0.4 to 4 by steps of 0.2 ; 0.4 (± 50 %)
Ventricular amplitude (V)
(2)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
Ventricular pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
(1) Values are measured using a positive and negative triangular signal of 2/13 ms.
(2) The correlation between the programmed amplitudes, the stored amplitudes and the mid-pulse
delivered amplitudes under a 500 ohm load are given in the following table:
Programmed ampl. (V) 1 1.5 2 2.5 3 3.5
Mid-pulse delivered
ampl. (V)
0.97 1.39 1.79 2.35 2.84 3.25
Stored amplitude (V) 1.14 1.63 2.1 2.76 3.33 3.82
Programmed ampl. (V) 4 4.5 5 6
Mid-pulse delivered ampl.
(V)
3.58 4.23 4.47 5.37
Stored amplitude (V) 4.2 4.96 5.25 6.3
46
SORIN – INTENSIA VR 124 – U153A