User's Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Declaration of conformity
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Lead connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Physical characteristics
- 19. Electrical characteristics
- 20. Programmable parameters
- 21. Non programmable parameters
- 22. Limited warranty
- 23. Patents
- 24. Explanation of symbols
20. PROGRAMMABLE PARAMETERS
3.5 3,65 3,09
4 4,2 3,55
4,5 4,69 3,97
5 5,25 4,44
6 6,3 5,33
Post-shock mode Values
Mode OFF-VVI
Duration 10s-20s-30s-1min-2min-3min-4min-5min
Basic rate (ppm) From 50 to 90 by steps of 5 ; 60 (± 4 %)
V amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
V pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
Sensitivity margins Values
Ventricular post pacing margin (mV) From 0 to 2 by steps of 0.2 ; 0.8
Response to noise Values
Automatic sensitivity on noise ON-OFF
V pacing on noise ON-OFF
48
SORIN – PLATINIUM VR – U461A