16. MAIN FUNCTIONS
16. MAIN FUNCTIONS
16.1. AUTOMATIC LEAD MEASUREMENTS
Automatic pacing lead impedance measurement:
A lead impedance measurement is automatically performed on the lead every 6 hours.The
daily mean impedance is stored.
Automatic coil impedance measurement:
A continuity measurement is automatically performed on defibrillation coil(s) once per day.
The continuity is stored for each coil.
Automatic sensing measurement:
The amplitude of R waves are automatically measured at each cycle. Every 8.5 minutes, the
amplitude of the last 8 R detections are averaged and stored.
16.2. VENTRICULAR TACHYARRHYTHMIA MANAGEMENT
Arrhythmia discrimination algorithm Stability/Acceleration and
Stability+/Acceleration:
Stability/Acceleration is the algorithm used to discriminate sinus tachycardias (ST) and
supraventricular tachycardias (SVT) from ventricular tachycardias (VT).
Stability+/Acceleration is based on the Stability/Acceleration algorithm but additionally takes
into account the “AFdetect” discrimination criteria: the occurrence of a ‘long ventricular cycle’
characteristic for AF patients which is an additional arrhythmia classification criterion to
improve identification of atrial fibrillation and avoid inappropriate shocks.
Fast VT treatment:
Applies detection criteria on fast ventricular tachycardia that are different from those of the
VT zone, as well as different therapies. The fast VT zone is included in the VF zone: its
lower limit is determined by the programmed value for the VF zone and its upper limit by the
programmed value for the fast VT zone.
Polarity alternation on Max shock:
Reverses the programmed polarity of every second shock set at maximum energy. The
number, type, and energy of shocks is independently programmable by detection zone.
Defibrillation threshold (DFT):
Be aware that the changes in the patient’s condition, drug regimen, and other factors may
change the defibrillation threshold (DFT) which may result in non-conversion of the
arrhythmia post-operatively. Successful conversion of ventricular fibrillation or ventricular
tachycardia during arrhythmia conversion testing is no assurance that conversion will occur
post-operatively.
36
SORIN PLATINIUM VR U461A