13. DEVICE DESCRIPTION
Coil impedance,
Capacitor charge time,
Pacing threshold tests.
Data storage:
Therapy History Report,
Statistics (pace/sense, therapy, shocks, and battery voltage),
Up to 16 complete Holter records with event logs, marker channel notation, and
electrogram records.
The PLATINIUM VR 1210 connector has three ports:
ventricular bipolar pace/sense and
two ports for RV and SVC defibrillation coils.
The PLATINIUM VR 1240 connector head has one port: ventricular bipolar pace/sense and
RV & SVC defibrillation coils.
The pace/sense port is compatible with the IS-1 standard and both defibrillation ports are
compatible with the DF-1 standard.
The defibrillation port is compatible with the DF-1 standard.
Distal lead terminal connections are secured with set-screws accessed via self-sealing
silicone plugs. All lead connections pass through the header into the device via
feedthroughs.
Programming System:
The Sorin programmer is used in conjunction with specific programmer software to
interrogate and program the implanted device at implant and during patient follow-up
procedures.
Remote Monitoring:
PLATINIUM VR is also equipped with the RF wireless technology which enables to remotely
monitor the patients who have the Sorin SMARTVIEW Monitor installed at home.
SORIN PLATINIUM VR U461A
27