User's Manual
Table Of Contents
- 1. Introduction
- 2. Definitions
- 3. Spider SAS Overview
- 4. Precautions for use and maintenance
- 5. Cleaning
- 6. User profile
- 7. General description
- 8. Patient Use
- 9. How to pair Spider SAS (Windows)
- 10. Starting a recording with HookUp RS
- 11. Starting recording: anonymous mode
- 12. Stopping a recording
- 13. Software, hardware and accessories
- 14. Technical specifications
- 15. Appendix 1: Patient recommendations
- 16. Appendix 2: electromagnetic compatibility
- 16.1. Guidance and manufacturer's declaration
- 16.2. Guidance and manufacturer's declaration: cable length
- 16.3. Guidance and manufacturer's declaration: electromagnetic emissions
- 16.4. Guidance and manufacturer's declaration: magnetic and electromagnetic immunity
- 16.5. Guidance and manufacturer's declaration: electromagnetic immunity, portable RF equipment
- 16.6. Spider SAS and mobiles (recommendations)
A new Lithium battery ensures 1 day 2 channel ECG recording including 2 nights of 8 hour
nocturnal respiratory polygraph.
NOTE: Place a new battery for each new recording in order to ensure the duration
announced above.
Internal memory battery
Spider SAS contains an internal battery recharged by the device’s main AAA battery. This
memory battery ensures the maintenance of the system’s date and time and other
parameters in the absence of the main battery for about 30 days.
NOTE: If the recorder has not been used for more than 1 month, it is recommended to
remove the AAA battery to avoid electrolyte leak (See § 4.1). Consequently, if the device is
stored for a very long period of time, the internal battery will run flat and therefore the
recorder will not start when re-using. It is then necessary to insert a new battery in the
recorder to recharge the internal memory battery before use. Depending on the level of
discharge of the internal battery, it is necessary to wait from a few minutes up to a few hours
to be able to start Spider SAS. The date and time of the recorder should be reset
appropriately by using HookUp RS before starting a new recording.
7.8.2.
7. GENERAL DESCRIPTION
SPIDER SAS – PHYSIO HUB – HOOKUP RS – U908F 23