User's Manual

ECG signal Quality
Make sure that the patient’s skin has been properly cleaned and dried before electrodes are
positioned. The upper channel displays lead A (red and white electrode), the lower channel
displays lead B (brown and white electrode). If the QRS amplitude on the ECG is lower than
0.5 mV (), you should check the positioning and the quality of the electrodes. Global quality
of ECG acquisition is shown independently for every channel (ECG 1 and ECG 2).
The bar graph color is fully blue when the quality of impedance is good enough ().
NOTE: If the amplitude of the signal is not sufficient, do not hesitate to change the position
of the electrode (red for channel A or brown for channel B). Cursor () is only for display
and will not modify the measured amplitude of the ECG.
Other Parameters
Fill out the patient details. All boxes are ticked when information is sent and received by
Spider SAS.
10.3.1.
10.3.2.
10. STARTING A RECORDING WITH HOOKUP RS
SPIDER SAS – PHYSIO HUB – HOOKUP RS – U908F 39