14.2. ELECTROMAGNETIC EMISSIONS
The Smart Spot is intended for use in the electromagnetic environment specified below. The customer or the
user of the Smart Spot should ensure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environment - guidance
Electromagnetic radiation
disturbance (radiated
EMISSIONS) CISPR 11
Group 1 The Smart Spot uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference to nearby electronic
equipment.
Disturbing voltage at supply
terminals conducted
emissions CISPR 11
Class B
The Smart Spot is suitable for use in home health care
environment and a professional health care establishment.
Harmonic current
emissions IEC 61000-3-2
Class A
Voltage variations, voltage
fluctuations and flicker
IEC 61000-3-3
Complies
14.3. ELECTROMAGNETIC IMMUNITY
The Smart Spot is intended for use in the electromagnetic environment specified below. The customer or the
user of the Smart Spot should ensure that it is used in such an environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Home health care environment and Professional health care
establishment.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for
power supply
lines
± 2 kV for
power supply
lines
Home health care environment and Professional health care
establishment.
± 1 kV for
input/output
lines
± 1 kV for
input/output
lines
Surge
IEC 61000-4-5
± 1 kV between
phases
± 1 kV between
phases
Home health care environment and Professional health care
establishment.
± 2 kV between
line(s) to earth
± 2 kV between
line(s) to earth
Magnetic field at
industrial rated
frequency
(IEC61000-4-8)
30 A/m 30 A/m Home health care environment and Professional health care
establishment.
Voltage dips
(IEC 61000-4-11)
0% U
T
For 0.5 cycle
A 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0% U
T
for 1
cycle
and
70% U
T
for
25 cycles,
50Hz
30 cycles, 60
Hz,
Monophase: 0°
0% U
T
For 0.5 cycle
A 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0% U
T
for 1
cycle
and
70% U
T
for
25 cycles,
50Hz
30 cycles, 60
Hz,
Monophase: 0°
Mains power quality should be that of a Home health care
environment or Professional health care establishment.
If operation of the system requires continued use during
power cuts, it is recommended that the medical device be
powered by a separate power source (UPS, etc.).
Smart Spot – UA10275A 9