User's Manual

ENGLISH 11
4.3. MEANING OF THE LIGHTS AND BUTTONS
Light or button State Colour Meaning
Status light OFF No colour The SMARTVIEW monitor is not
plugged in properly. Please check the
connections. If the status light remains
off, please contact our toll-free technical
assistance.
ON Constant
amber
You have just connected the
SMARTVIEW monitor to the wall
socket. The SMARTVIEW monitor is
setting up. Please wait. This process
takes less than 5 minutes.
Constant
green
The SMARTVIEW monitor is operational.
Patient initiated
transfer button
Flashing Green The SMARTVIEW monitor is gathering
data. Stay as close as possible to the
SMARTVIEW monitor and don’t
disconnect it.
Red The SMARTVIEW monitor could not
gather the information stored in your
implanted defibrillator. Move closer to the
SMARTVIEW monitor and restart the
process. If the button turns red again,
please contact our toll-free technical
assistance.
ON Green You pushed the button to request a Patient
Initiated Transfer transmission. The
SMARTVIEW monitor is checking that the
function is active before performing the
transmission.
Progress lights Progress Green The SMARTVIEW monitor is gathering
information from your implanted
defibrillator. Stay as close as possible
to the SMARTVIEW monitor and don’t
disconnect it.