Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic
environment -
guidance
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
<5 % UT
(>95 % dip in
UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in
UT)
for 5 s
Mains power quality
should be that of a
typical home
environment. If the
user of the
SMARTVIEW monitor
requires continued
operation during
power mains
interruptions, it is
recommended that
the SMARTVIEW
monitor be powered
from an
uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m Power frequency
magnetic fields should
be at levels
characteristic of a
typical bedroom in a
typical home
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
The SMARTVIEW monitor is intended for use in the electromagnetic environment specified
below. The customer or the user of the SMARTVIEW monitor should assure that it is used in
such an environment.
Portable and mobile RF communications equipment should be used no closer to the
SMARTVIEW monitor, including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.