User Manual

6. WARNINGS AND PRECAUTIONS
SORIN ORCHESTRA PLUS LINK U386A
7
6. WARNINGS AND PRECAUTIONS
6.1 SAFETY PRECAUTIONS
Alteration Your ORCHESTRA PLUS LINK and supplied accessories shall not be
modified, altered, or changed in any way without signed written permission from Sorin.
Unauthorized modification may void the equipment authorization from the certified body
(compliance to standards) and compromise system performance. Use only Sorin supplied
accessories.
Environmental factors - To reduce the risk of system failure, fire, personal injury, do not
expose this appliance to humidity, moisture, extreme temperatures or electrical storms. If the
ORCHESTRA PLUS LINK is damaged due to a mechanical shock or liquid, do not plug it in
again and call your sales representative.
Keep the device dry Do not spill or drip liquids on it.
6.2 PRECAUTIONS FOR SAFE OPERATION
Location Preferably the ORCHESTRA PLUS LINK should be placed in a stable position
as close as possible to the patient.
Power supply The ORCHESTRA PLUS LINK shall be connected exclusively to one of
Sorin Orchestra Plus USB ports. Refer to Orchestra Plus user manual for location of ports E
& C (dual ports).
Power supply protection ORCHESTRA PLUS LINK USB cable shall be routed so it is not
likely to be walked on or pinched by items placed on or against it. In addition, USB cable
extensions shall not be used to connect ORCHESTRA PLUS LINK to Orchestra Plus.
Power supply connection ORCHESTRA PLUS LINK USB cable shall be plugged into
Orchestra PLUS before interrogating the implantable medical device. It will not be possible
to establish an RF communication with the implantable medical device if the interrogation
starts before the ORCHESTRA PLUS LINK has been connected.
Power supply disconnection You can only remove power from ORCHESTRA PLUS
LINK by disconnecting its USB cable from Orchestra Plus. Always carefully disconnect
ORCHESTRA PLUS LINK by pulling on the USB plug and not on the cable.
Do not remove the USB cable during the Orchestra Plus programming session of the
implantable medical device.
Interferences ORCHESTRA PLUS LINK can be subject to disturbances from other
products in the vicinity which emit electromagnetic interference (see Troubleshooting section
10). Keep the space clear between the ORCHESTRA PLUS LINK and the implantable
medical device.