12. TROUBLESHOOTING
Start-up problems
If the ORCHESTRA PLUS LINK functioning status LED is not green, verify USB connection
with Orchestra Plus. If the problem persists, call your sales representative.
Unstable or no RF communication with implantable medical device
To recover stable communication:
- Verify USB connection with Orchestra Plus
- Move the inductive programming head away from the implantable medical device
during the implantation or follow-up session, when the RF session is initiated.
- Remove any objects between ORCHESTRA PLUS LINK and implantable medical
device
- Re-position (5 - 10 cm) or rotate the ORCHESTRA PLUS LINK
- Move the ORCHESTRA PLUS LINK closer to the implantable medical device
- If none of the above works, switch to inductive telemetry