User's Manual
Table Of Contents
- 1. Warnings and precautions
- 2. Warranty conditions
- 3. The programmer system
- 4. The programmer tablet
- 5. The programmer software
- 6. The docking station
- 7. The inductive programming head
- 8. The radio frequency (RF) link
- 9. The ECG
- 10. The printer
- 11. Maintenance
- 12. Guidance and manufacturer's declaration
- 12.1. Radio equipment emission
- 12.2. Recommended separation distances according to the maximum output power of the communications equipment
- 12.3. Cable length
- 12.4. Electromagnetic emissions
- 12.5. Magnetic and Electromagnetic Immunity
- 12.6. Electromagnetic Immunity, Portable Radio Frequency Equipment
- 12.7. Raw materials
- 13. Technical data
- 13.1. Conformity of the programmer
- 13.2. US Federal Communications Commissions (FCC) and Industry Canada
- 13.3. Essential performances
- 13.4. Power supply
- 13.5. Dimensions programmer tablet
- 13.6. Dimensions docking station
- 13.7. Dimensions inductive programming head
- 13.8. Use constraints
- 13.9. Storage constraints
- 13.10. System specifications of the programmer
- 14. Explanation of symbols
- 15. Glossary
Cables and
accessories
Maximum length
Test type
In compliance with
Immunity to conducted
disturbances, induced by
radio-frequency fields
IEC61000-4-6
Radiated immunity -
Magnetic fields
IEC61000-4-8
Voltage dips, short
interruptions and voltage
variation immunity
IEC 61000-4-11
ELECTROMAGNETIC EMISSIONS
The programmer is intended for use in the electromagnetic environment specified below. The
customer or the user of the programmer should ensure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environment - guidance
Electromagnetic radi-
ation disturbance (ra-
diated EMISSIONS)
CISPR 11
Group 1 The programmer uses RF energy only for its own func-
tion. Its RF emissions are very low and are not likely to
cause any interference to nearby electronic equipment.
Disturbing voltage at
supply terminals con-
ducted emissions
CISPR 11
Class B The programmer is suitable for use in a professional
health care establishment.
Harmonic current
EMISSIONS IEC
61000-3-2
Class A
Voltage variations,
voltage fluctuations
and flicker IEC
61000-3-3
Complies
MAGNETIC AND ELECTROMAGNETIC IMMUNITY
The programmer is intended for use in the electromagnetic environment specified below. The
customer or the user of the programmer should ensure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
± 8 kV con-
tact
± 15 kV air
± 8 kV con-
tact
± 15 kV air
Professional health care establishment.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for
power sup-
ply lines
± 2 kV for
power sup-
ply lines
Professional health care establishment.
± 1 kV for
input/output
lines
± 1 kV for in-
put/output
lines
Surge
IEC 61000-4-5
± 1 kV be-
tween phas-
es
± 1 kV be-
tween phas-
es
Professional health care establishment.
12.4.
12.5.
12. GUIDANCE AND MANUFACTURER'S DECLARATION
SmartTouch – UA10412A 27