User's Manual
Table Of Contents
- 1. Warnings and precautions
- 2. Warranty conditions
- 3. The programmer system
- 4. The programmer tablet
- 5. The programmer software
- 6. The docking station
- 7. The inductive programming head
- 8. The radio frequency (RF) link
- 9. The ECG
- 10. The printer
- 11. Maintenance
- 12. Guidance and manufacturer's declaration
- 12.1. Radio equipment emission
- 12.2. Recommended separation distances according to the maximum output power of the communications equipment
- 12.3. Cable length
- 12.4. Electromagnetic emissions
- 12.5. Magnetic and Electromagnetic Immunity
- 12.6. Electromagnetic Immunity, Portable Radio Frequency Equipment
- 12.7. Raw materials
- 13. Technical data
- 13.1. Conformity of the programmer
- 13.2. US Federal Communications Commissions (FCC) and Industry Canada
- 13.3. Essential performances
- 13.4. Power supply
- 13.5. Dimensions programmer tablet
- 13.6. Dimensions docking station
- 13.7. Dimensions inductive programming head
- 13.8. Use constraints
- 13.9. Storage constraints
- 13.10. System specifications of the programmer
- 14. Explanation of symbols
- 15. Glossary
12. GUIDANCE AND MANUFACTURER'S DECLARATION
NOTE: For the guidance and manufacturer’s declaration of the Radio Frequency (RF) Link
accessory, please refer to the associated user manual.
RADIO EQUIPMENT EMISSION
Radio Equipment
Transmitter Frequency Bands/
Maximal Power
Receiver Frequency
Bands
Orchestra Plus Link
[2400; 2483] MHz; 10 mW (+10 dBm) [402 – 405] MHz
[402 – 405] MHz; 25 µW (-16 dBm)
SmartTouch Tablet
Bluetooth Low Energy and
Bluetooth
[2400; 2483.5] MHz; ≤10 mW (+10
dBm)
[2400 – 2483.5] MHz
Inductive Telemetry CPR3H
Slow mode: frequency ~30kHz modu-
lation pulsed,
12.4 dBµA/m at 3m
Fast mode: frequency ~70 kHz modu-
lation pulsed,
17.4 dBµA/m at 3m
8kHz, 16kHz
Note: Values provided are in accordance with Directive 2014/53/EU.
RECOMMENDED SEPARATION DISTANCES ACCORDING TO THE
MAXIMUM OUTPUT POWER OF THE COMMUNICATIONS
EQUIPMENT
The tablet programmer is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled.
Portable RF communication devices (including peripherals such as antenna cables and
external antennas) should not be closer to 30 cm (12 inches) from any part of the
programmer. Otherwise, the performance of these devices may be impaired.
CABLE LENGTH
Cables and
accessories
Maximum length
Test type
In compliance with
Cables/Cords < 3m
RF emission CISPR 11, Class B
Harmonic current emission IEC61000-3-2
Voltage fluctuation and
flickers
IEC61000-3-3
Electrostatic discharge immunity IEC61000-4-2
Radiated immunity –
Electromagnetic fields
IEC61000-4-3
Electrical fast transient/
burst immunity
IEC61000-4-4
Surge immunity IEC61000-4-5
12.1.
12.2.
12.3.
12. GUIDANCE AND MANUFACTURER'S DECLARATION
26 SmartTouch – UA10412A