User's Manual
Table Of Contents
- 1. Warnings and precautions
- 2. Warranty conditions
- 3. The programmer system
- 4. The programmer tablet
- 5. The programmer software
- 6. The docking station
- 7. The inductive programming head
- 8. The radio frequency (RF) link
- 9. The ECG
- 10. The printer
- 11. Maintenance
- 12. Guidance and manufacturer's declaration
- 12.1. Radio equipment emission
- 12.2. Recommended separation distances according to the maximum output power of the communications equipment
- 12.3. Cable length
- 12.4. Electromagnetic emissions
- 12.5. Magnetic and Electromagnetic Immunity
- 12.6. Electromagnetic Immunity, Portable Radio Frequency Equipment
- 12.7. Raw materials
- 13. Technical data
- 13.1. Conformity of the programmer
- 13.2. US Federal Communications Commissions (FCC) and Industry Canada
- 13.3. Essential performances
- 13.4. Power supply
- 13.5. Dimensions programmer tablet
- 13.6. Dimensions docking station
- 13.7. Dimensions inductive programming head
- 13.8. Use constraints
- 13.9. Storage constraints
- 13.10. System specifications of the programmer
- 14. Explanation of symbols
- 15. Glossary
8. Remove the USB key and press “OK” in order to confirm that the installation was
successful.
9. The programmer tablet will shut down automatically and will be ready for use when
restarted.
STORING THE PATIENT DATA
Patient data is stored on the programmer tablet hard drive.
We recommend regular back-up of the patient data stored on the programmer tablet hard
drive.
In case the programmer has to be retrieved from the field all patient data must be deleted
from the programmer hard drive. Patient data can be saved on an external storage medium
prior to deletion.
1. Connect a USB flash drive to the USB port at the left side of the programmer tablet or
to the USB port of the docking station, in case the tablet is placed in the docking during
usage.
2. Select the files required and press the “Export” button to transfer the files to the chosen
external storage medium (see Figure - Export files screen below).
― In order to protect the patient data from unauthorized use, the patient data files will be
automatically anonymized when exported to an external storage medium.
― During this automated file anonymization the patient’s name will be deleted in the
exported file.
― CAUTION: In addition to deleting the patient’s name, the patient’s birthdate will be
anonymized by changing the original day of birth to the “15th” of the same month.
However, the original month and year of patient’s birth shall be maintained.
UPGRADING THE PROGRAMMER SOFTWARE
Before starting a software upgrade on the programmer, plug the tablet to the power supply to
prevent the tablet shutting down during the software upgrading process.
5.3.
5.4.
5. THE PROGRAMMER SOFTWARE
SmartTouch – UA10412A 15