User's Manual

US-ENGLISH 69
Ventricular arrhythmia
prevention
Values
Atrial pacing on PVC
Yes-No
Post extrasystolic pause suppression
Yes-No
Acceleration on PVC
Yes-No
Max accelerated rate (min-1)
From 60 to 145 by steps of 5 ; 100
Post-shock mode
Values
Mode
OFF-VVI-DDI-DDD
Duration
10s-20s-30s-1min-2min-3min-4min-5min
Basic rate (min-1)
From 50 to 90 by steps of 5 ; 60 4 %)
Rest AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-
115-125-135-140-150-155-165-170-180-
190-195-205-210-220-225-235-250
19 ms)
Exercise AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-
115-125-135-140-150-155-165-170-180-
190-195-205-210-220-225-235-250
19 ms)
AVD Paced/Sensed Offset (ms)
0-10-15-25-30-40-45-55-65-70-80-85-95-
100-110-115-125 (± 1 ms)
A amplitude (V)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
A pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
10 %)
V amplitude (V)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
V pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
10 %)