20. PROGRAMMABLE PARAMETERS
measured at 37 °C under a 500 ohm load
Legend:
Value in bold: as shipped” value
Underlined value: nominal value
20.1. ANTIBRADYCARDIA PACING
Basic parameters
Values
Mode
VVI-VVIR-DDD-DDDR-DDD/DDIR-DDI-
DDIR-SafeR (AAI <=> DDD)-SafeR-R
(AAIR <=> DDDR)
Basic rate (min-1) (1)
From 30 to 90 by steps of 5 ; 60 4 %)
Maximum rate (min-1)
From 100 to 145 by steps of 5 ; 120 6 %)
Rate hysteresis (%)
0-5-10-20-35 (± 18 ms)
Rest AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-115-
125-135-140-150-155-165-170-180-190-
195-205-210-220-225-235-250 (± 19 ms)
Exercise AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-115-
125-135-140-150-155-165-170-180-190-
195-205-210-220-225-235-250 (± 19 ms)
AVD Paced/Sensed Offset (ms)
0-10-15-25-30-40-45-55-65-70-80-85-95-
100-110-115-125 (± 1 ms)
(1) The corresponding periods are (in ms): 2000-1714-1500-1333-1200-1091-
1000-923-857-800-750-706-667 ms.