User's Manual
46 – US-ENGLISH
15. SPECIAL MODES
15.1. SAFETY MODE (NOMINAL VALUES)
Nominal values may be rapidly restored by pressing the following
button on the programming head or programmer keyboard:
or via the "Emergency" button on the SMARTVIEW screen.
In safety mode, the defibrillator operates with the parameters
underlined in the table of programmable parameters.
15.2. MAGNET MODE
When the magnet is applied:
antiarrhythmia functions are inhibited (detection of rhythm
disturbances, charging, and therapy),
hysteresis and AVD paced/sensed offset are set to 0,
pacing amplitude is set to 6 V,
pulse width is set to maximum,
pacing rate is set to the magnet rate,
the following functions are disabled: ventricular arryhtmia
prevention, Mode Switch, Anti-PMT, Smoothing, Rate Response.
When the magnet is removed:
the sensor rate is forced to the basic rate,
arrhythmia detection algorithms and sequential therapies are
reinitialized,
therapies start with the least aggressive program for each area.
The other parameters remain at their programmed value.