User's Manual

US-ENGLISH 43
14.9. LEAD CONNECTION
Implant the ventricular lead, then the atrial lead.
Each lead must be connected to the corresponding connector port.
The position of each connector is indicated on the casing.
CAUTION: Tighten only the distal inserts.
To connect each lead, proceed as follows:
1. Clean the lead terminal pins thoroughly, if necessary (device
replacement).
2. Lubricate the lead terminal pins with sterile water, if necessary.
3. Do not insert a lead connector pin into the connector block
without first visually verifying that the lead port is not filled with
any obstacle.
4. Insert the screwdriver into the pre-inserted screw socket of the
appropriate port (in order to allow excess air to bleed out and to
make the insertion of the lead pin easier).
5. Insert the lead pin all the way into the port (check that the pin
protrudes beyond the distal insert).
6. Tighten, check the tightness and ensure the lead pin still
protrudes beyond the distal insert, and did not move.