User's Manual

38 US-ENGLISH
Up to 14 complete Holter records with event logs, marker
channel notation, and electrogram records.
The connector head has four ports: atrial bipolar pace/sense,
ventricular bipolar pace/sense and two ports for RV and SVC
defibrillation coils. Both pace/sense ports are compatible with the IS-
1 standard and both defibrillation ports are compatible with the DF-1
standard. Distal lead terminal connections are secured with set-
screws accessed via self-sealing silicone plugs. All lead connections
pass through the header into the device via feedthroughs.
Programming System: The Sorin CRM programmer is used in
conjunction with specific programmer software to interrogate and program
the implanted device at implant and during patient follow-up procedures.
Remote Monitoring: The PARADYM RF DR 9550 is also equipped
with the RF wireless technology which enables to remotely monitor
the patients who have the Sorin CRM SMARTVIEW Monitor installed
at home.
14. IMPLANT PROCEDURE
14.1. NECESSARY EQUIPMENT
Implantation of PARADYM RF DR 9550 requires the following
equipment:
Sorin CRM dedicated programmer, equipped with the
SMARTVIEW software interface and with the programming head,
pacing system analyser, as well as its sterile connecting cables,
to evaluate the pacing and sensing thresholds,
a complete set of leads with corresponding introducers,
physiological signal monitor capable of displaying simultaneously
the surface ECG and arterial pressure,
an external defibrillator with sterile external paddles,
sterile cover for the telemetry head.