20. PROGRAMMABLE PARAMETERS
Post-shock mode Values
Mode OFF-VVI-DDI-DDD
Duration 10s-20s-30s-1min-2min-3min-4min-5min
Basic rate (ppm) From 50 to 90 by steps of 5 ; 60 (± 4 %)
Rest AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-
155-165-170-180-190-195-205-210-220-225-235-250
(± 19 ms)
Exercise AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-
155-165-170-180-190-195-205-210-220-225-235-250
(± 19 ms)
AVD Paced/Sensed Offset (ms) 0-10-15-25-30-40-45-55-65-70-80-85-95-100-110-115-125
(± 1 ms)
A amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
A pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
V amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
V pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
Refractory periods Values
Atrial refractory period post
ventricular sensing (ms)
45-65-80-95-110-125-140-155 (± 16 ms)
Atrial refractory period post
ventricular pacing (ms)
80-95-110-125-140-155 (± 4 ms)
Sensitivity margins Values
Atrial post pacing/sensing margin
(mV)
From 0 to 1 by steps of 0.2 ; 0.4
Ventricular post pacing margin (mV) From 0 to 2 by steps of 0.2 ; 0.8
Response to noise Values
Automatic sensitivity on noise ON-OFF
V pacing on noise ON-OFF
SafeR parameters Values
AVB I switch Rest+Exercise-Exercise
Long PR: max (ms) From 200 to 500 by steps of 50 ; 450
Long PR: min (ms) From 200 to 500 by steps of 50 ; 250
Max. pause (s) 2-3-4
SORIN PLATINIUM DR U460A