19. ELECTRICAL CHARACTERISTICS
Longevity projection at 700 Ω pacing impedance:
Mode DDD DDD DDD DDD DDD SafeR -
A pacing (%) 100 100 15 15 15 30 0
V pacing (%) 100 100 1 15 1 6 0
Pacing amplitude (V) 3.5 3.5 3.5 4.5 2.5 2.5 -
Pulse width (ms) 0.35 0.35 0.35 0.50 0.35 0.35 -
Sensor OFF ON OFF OFF OFF OFF OFF
Max shocks (42J) per year 4 4 4 4 0 4 4
Longevity (years) 9.5 9.3 13.3 12.0 17.1 13.2 13.7
1h of additional RF programming session reduces the device longevity from 1 to 2 weeks
depending on the device functioning mode (no pacing, 100% pacing). The mean longevity
as a function of shocks delivered at maximum energy, with and without pacing, is as follows:
The mean longevity as a function of yearly remote follow-ups
1
, with and without pacing, is as
follows:
1. An excessive number of remote follow-ups can have a non-negligible impact on device
longevity.
52
SORIN PLATINIUM DR U460A