User's Manual

Table Of Contents
22. NON PROGRAMMABLE PARAMETERS
Interval Values Nominal value "As shipped"
value
Committed period
95 ms 5 ms)
Atrial refractory pe-
riods
Values Nominal value "As shipped"
value
Post atrial sensing
47 ms 16 ms)
Post atrial pacing
109 ms 4 ms)
Ventricular refrac-
tory periods
Values Nominal value "As shipped"
value
Post ventricular
sensing
95 ms 16 ms)
Post ventricular pac-
ing
220 ms 4 ms)
Post atrial pacing
(blanking)
16 ms 3 ms)
Therapies Values Nominal value "As shipped"
value
Waveform
(1)
Constant tilt (50% - 50%)
Stored energy for the
Max. shock
42 J
Pacing amplitude
during ATP therapies
7 V (Actual value at 300 ms: 5.3 V )
Anti-PMT protection
Termin
(1) The device has 50% tilt in each phase thus delivers 94% of stored energy. Each phase is limited
to 10 ms duration.
22. NON PROGRAMMABLE PARAMETERS
SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A 69