User's Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Declaration of conformity
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Leads connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Supplemental Information
- 19. Physical characteristics
- 20. Electrical characteristics
- 21. Programmable parameters
- 22. Non programmable parameters
- 23. Limited warranty
- 24. Patents
- 25. Explanation of symbols
22. NON PROGRAMMABLE PARAMETERS
Interval Values Nominal value "As shipped"
value
Committed period
95 ms (± 5 ms)
Atrial refractory pe-
riods
Values Nominal value "As shipped"
value
Post atrial sensing
47 ms (± 16 ms)
Post atrial pacing
109 ms (± 4 ms)
Ventricular refrac-
tory periods
Values Nominal value "As shipped"
value
Post ventricular
sensing
95 ms (± 16 ms)
Post ventricular pac-
ing
220 ms (± 4 ms)
Post atrial pacing
(blanking)
16 ms (± 3 ms)
Therapies Values Nominal value "As shipped"
value
Waveform
(1)
Constant tilt (50% - 50%)
Stored energy for the
Max. shock
42 J
Pacing amplitude
during ATP therapies
7 V (Actual value at 300 ms: 5.3 V )
Anti-PMT protection
Termin
(1) The device has 50% tilt in each phase thus delivers 94% of stored energy. Each phase is limited
to 10 ms duration.
22. NON PROGRAMMABLE PARAMETERS
SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A 69