up, the information is managed by a warning (pop-up message) to immediately notify the
user. For example, the following types of events can trigger a warning or an alert: technical
problem during a shock, pacing lead impedance or coil impedance measurements out-of-
range, battery depletion, etc. The Remote tab presents an overview of all the alerts managed
by the device.
General parameters Values Nominal value "As shipped"
value
RF communication
(1)
ON-OFF ON OFF
Alerts
(1)
ON-OFF ON ON
(1) RF and Remote alerts are turned on automatically if Shocks are programmed ON.
When Alerts are programmed "On", the following System Alerts are automatically activated:
“Battery depletion – RRT”
“Device reset”
“Excessive charge time (>25s)”
“System integrity”
Lead Alerts Values Nominal value "As shipped"
value
Abnormal A lead im-
pedance
ON-OFF ON ON
Abnormal A lead low
limit (Ohm)
200-250-300-350-400-450-500 200 200
Abnormal A lead
high limit (Ohm)
1500-1750-2000-2500-3000 3000 3000
Abnormal RV lead
impedance
ON-OFF ON ON
Abnormal RV lead
low limit (Ohm)
200-250-300-350-400-450-500 200 200
Abnormal RV lead
high limit (Ohm)
1500-1750-2000-2500-3000 3000 3000
Abnormal LV lead
impedance
ON-OFF ON ON
Abnormal LV lead
low limit (Ohm)
200-250-300-350-400-450-500 200 200
Abnormal LV lead
high limit (Ohm)
1500-1750-2000-2500-3000 3000 3000
Abnormal RV coil im-
pedance
ON-OFF ON ON
Abnormal SVC coil
impedance
ON-OFF ON ON
Abnormal Shock im-
pedance
(1)
ON-OFF ON ON
(1) Normal impedance range [20 Ohm-200 Ohm]