User's Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Declaration of conformity
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Leads connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Supplemental Information
- 19. Physical characteristics
- 20. Electrical characteristics
- 21. Programmable parameters
- 22. Non programmable parameters
- 23. Limited warranty
- 24. Patents
- 25. Explanation of symbols
up, the information is managed by a warning (pop-up message) to immediately notify the
user. For example, the following types of events can trigger a warning or an alert: technical
problem during a shock, pacing lead impedance or coil impedance measurements out-of-
range, battery depletion, etc. The Remote tab presents an overview of all the alerts managed
by the device.
General parameters Values Nominal value "As shipped"
value
RF communication
(1)
ON-OFF ON OFF
Alerts
(1)
ON-OFF ON ON
(1) RF and Remote alerts are turned on automatically if Shocks are programmed ON.
When Alerts are programmed "On", the following System Alerts are automatically activated:
― “Battery depletion – RRT”
― “Device reset”
― “Excessive charge time (>25s)”
― “System integrity”
Lead Alerts Values Nominal value "As shipped"
value
Abnormal A lead im-
pedance
ON-OFF ON ON
Abnormal A lead low
limit (Ohm)
200-250-300-350-400-450-500 200 200
Abnormal A lead
high limit (Ohm)
1500-1750-2000-2500-3000 3000 3000
Abnormal RV lead
impedance
ON-OFF ON ON
Abnormal RV lead
low limit (Ohm)
200-250-300-350-400-450-500 200 200
Abnormal RV lead
high limit (Ohm)
1500-1750-2000-2500-3000 3000 3000
Abnormal LV lead
impedance
ON-OFF ON ON
Abnormal LV lead
low limit (Ohm)
200-250-300-350-400-450-500 200 200
Abnormal LV lead
high limit (Ohm)
1500-1750-2000-2500-3000 3000 3000
Abnormal RV coil im-
pedance
ON-OFF ON ON
Abnormal SVC coil
impedance
ON-OFF ON ON
Abnormal Shock im-
pedance
(1)
ON-OFF ON ON
(1) Normal impedance range [20 Ohm-200 Ohm]
21. PROGRAMMABLE PARAMETERS
SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A 67