Pacing/Sensing Values Nominal value "As shipped"
value
LV amplitude – MP1
(V)
(1)
0.1-0.25-0.35 80 %)
0.5-0.6-0.75-0.85 30 %)
1-1.1-1.25-1.35-1.5-1.6-1.75-1.85-2-2.1-
2.25-2.35-2.5-2.6-2.75-2.85-3-3.1-3.25-
3.35-3.5-3.6-3.75-3.85-4-4.1-4.25-4.35-
4.5-4.6-4.75-4.85-5-5.1-5.25-5.35-5.5-
5.6-5.75-5.85-6-6.1-6.25-6.35-6.5-6.6-
6.75-6.85-7- 20 %)
5 3.5
LV pulse width –
MP1 (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1-1.1-
1.25-1.35-1.5-1.6-1.75-1.85-2 10 %)
0.35 0.35
LV pacing polarity –
MP1
[LV tip1-CAN]-[LV tip1-LV2]-
[LV tip1-RV ring]-[LV tip1-RV coil]-
[LV tip1-LV4]-[LV2-CAN]-[LV2- RV coil]-
[LV2 -LV4]
[LV tip1-LV2] [LV tip1-LV2]
LV amplitude – MP2
(V)
(1)
0.1-0.25-0.35 80 %)
0.5-0.6-0.75-0.85 30 %)
1-1.1-1.25-1.35-1.5-1.6-1.75-1.85-2-2.1-
2.25-2.35-2.5-2.6-2.75-2.85-3-3.1-3.25-
3.35-3.5-3.6-3.75-3.85-4-4.1-4.25-4.35-
4.5-4.6-4.75-4.85-5-5.1-5.25-5.35-5.5-
5.6-5.75-5.85-6-6.1-6.25-6.35-6.5-6.6-
6.75-6.85-7- 20 %)
5 3.5
LV amplitude – MP2
(ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1-1.1-
1.25-1.35-1.5-1.6-1.75-1.85-2 10 %)
0.35 0.35
LV pacing polarity –
MP2
[LV3-CAN]-[LV3-LV4]-[LV3-LV2]-
[LV3-RV ring]-[LV3-RVcoil]-[LV4-RV coil]
[LV3-LV4] [LV3-LV4]
(1) The correlation between the programmed amplitudes, the stored amplitudes and the mid-pulse
delivered amplitudes under a 500 ohm load are given in the table above.
Ventricular arrhyth-
mia algorithms
Values Nominal value "As shipped"
value
Atrial pacing on PVC
Yes-No No No
Post extrasystolic
pause suppression
Yes-No No No
Acceleration on PVC
Yes-No No No
Max acceleration
rate (ppm)
From 60 to 145 by steps of 5 100 100
21. PROGRAMMABLE PARAMETERS