User's Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Declaration of conformity
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Leads connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Supplemental Information
- 19. Physical characteristics
- 20. Electrical characteristics
- 21. Programmable parameters
- 22. Non programmable parameters
- 23. Limited warranty
- 24. Patents
- 25. Explanation of symbols
Programmed
amplitude
(V)
Stored
amplitude
(V)
Mid-pulse
delivered
amplitude
(V)
2.10* 2.21 1.87
2.25* 2.34 1.98
2.35* 2.47 2.09
2.50 2.60 2.20
2.60* 2.73 2.31
2.75* 2.87 2.42
2.85* 3.00 2.53
3.00 3.15 2.67
3.10* 3.26 2.76
3.25* 3.39 2.87
3.35* 3.52 2.98
3.50 3.65 3.09
3.60* 3.78 3.20
3.75* 3.91 3.31
3.85* 4,04 3.42
4.00 4.20 3.55
4.10* 4.30 3.64
4.25* 4.43 3.75
4.35* 4.56 3.86
4.50 4.69 3.97
Programmed
amplitude
(V)
Stored
amplitude
(V)
Mid-pulse
delivered
amplitude
(V)
4.60* 4.82 4.08
4.75* 4.95 4.19
4.85* 5.08 4.30
5.00 5.25 4.44
5.10* 5.34 4.52
5.25* 5.47 4.63
5.35* 5.60 4.74
5.50* 5.73 4.85
5.60* 5.86 4.96
5.75* 5.99 5.07
5.85* 6.12 5.18
6.00 6.30 5.33
6.10* 6.38 5.40
6.25* 6.51 5.51
6.35* 6.64 5.62
6.50* 6.77 5.73
6.60* 6.90 5.84
6.75* 7.03 5.95
6.85* 7.16 6.06
7.00* 7.35 6.22
* For left ventricular amplitude only.
If Multipoint LV Pacing (MP) is set to ON, LV pacing parameters (LV amplitude, LV pulse width
and LV pacing polarity) are replaced by the following parameters:
21. PROGRAMMABLE PARAMETERS
58 SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A