21. PROGRAMMABLE PARAMETERS
Measured at 37 °C under a 500 ohm load.
ANTIBRADYCARDIA PACING
Basic parameters Values Nominal value "As shipped"
value
Mode
VVI-VVIR-DDD-DDDR-DDD/DDIR-DDI-
DDIR-SafeR (AAI <=> DDD)-
SafeR-R (AAIR <=> DDDR)-DOO-VOO-
OOO
VVI DDD
Basic rate (ppm)
(1)
From 30 to 90 by steps of 5 (± 4 %) 60 60
Maximum rate (ppm)
From 100 to 145 by steps of 5 (± 6 %) 120 120
Rate hysteresis (%)
0-5-10-20-35 18 ms) 0 0
Rest AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-
115-125-135-140-150-155-165-170-180-
190-195-205-210-220-225-235-250
19 ms)
125 125
Exercise AV delay
(ms)
30-40-45-55-65-70-80-85-95-100-110-
115-125-135-140-150-155-165-170-180-
190-195-205-210-220-225-235-250
19 ms)
80 80
AVD Paced/Sensed
Offset (ms)
0-10-15-25-30-40-45-55-65-70-80-85-95
-100-110-115-125 1 ms)
65 65
(1) The corresponding periods are (in ms):
2000-1714-1500-1333-1200-1091-1000-923-857-800-750-706-667 ms.
Special features Values Nominal value "As shipped"
value
Rate smoothing
OFF-Very slow-Slow-Medium-Fast OFF OFF
Mode Switch
ON-OFF ON ON
Physical activity
Very low-Low-Medium-High-Very high Medium Medium
Exercise AV opt. rate
(ppm)
From 70 to 120 by steps of 5 90 90
CRT optimization
OFF-Monitor-AV-AV+VV OFF OFF
21.1.
21. PROGRAMMABLE PARAMETERS
56 SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A