User's Manual

Table Of Contents
Common mode rejection ratio:
16.6 Hz 50 Hz 60 Hz
Atrial channel ≥74 dB ≥74 dB ≥74 dB
Ventricular
channel
≥68 dB ≥68 dB ≥68 dB
For atrial sensitivity settings below 0.4mV, the ICD may detect noise lower than the level
specified in clause 27.5.1 of standard EN 45502-2-2 for frequencies below 200 Hz.
For ventricular sensitivity settings below 0.6mV, the ICD may detect noise lower than the level
specified in clause 27.5.1 of standard EN 45502-2-2 for frequencies below 200 Hz.
PROTECTION AGAINST SHORT-CIRCUITS
The defibrillator can undergo a short-circuit if the anode and cathode are not adequately
separated.
In this case, the shock is aborted and a warning will indicate that a short circuit (shock
impedance < 20 ohms) was detected during the last shock. The device may be damaged
compromising ability to provide shock therapy.
15.5.
15. SPECIAL MODES
40 SORIN – PLATINIUM 4LV SonR CRT-D 1844 – U906A