User's Manual

Table Of Contents
13. DEVICE DESCRIPTION
Programmer-controlled VT induction sequences,
Programmer-controlled VF inductions (30 Hz rapid pacing or shock on T),
Programmable electrogram vectors (A / RV coil-CAN / SVC coil-CAN /RV coil-SVC
coil / LV bip / LV tip-RV ring / A ring-CAN / RV tip-CAN / RV ring-CAN / LV tip-CAN /
LV ring-CAN) and RV EGM,
Real-time annotations displayed with the markers and indicating the majority rhythm,
Manual ATP sequences,
Manual shocks.
Rescue shock
Follow-up tests:
Pacing lead impedance,
Coil impedance,
Capacitor charge time,
Sensitivity test,
Pacing threshold tests.
Data storage:
Therapy History Report,
Statistics (pace/sense, therapy, shocks, and battery voltage),
Up to 16 complete Holter records with event logs, marker channel notation, and
electrogram records.
The PLATINIUM SonR CRT-D 1811 connector has five ports:
Atrial “SonR” port: performs atrial bipolar pace/sense if a conventional IS-1 lead is
connected as well as SonR capability if the SonR atrial lead is connected,
RV “IS-1” port: performs right ventricular bipolar pace/sense,
LV “IS-1” port: performs left ventricular bipolar pace,
RV “DF-1” port for RV defibrillation coil,
SVC “DF-1” port for SVC defibrillation coil.
The PLATINIUM SonR CRT-D 1841 connector head has three ports:
Atrial “SonR” port: performs atrial bipolar pace/sense if a conventional IS-1 lead is
connected as well as SonR capability if the SonR atrial lead is connected,
LV “IS-1” port: performs left ventricular bipolar pace,
RV “DF4” port: performs right ventricular bipolar pace/sense, port for RV/SVC
defibrillation coils.
Distal lead terminal connections are secured with set-screws accessed via self-sealing
silicone plugs. All lead connections pass through the header into the device via
feedthroughs.
Programming System:
The Sorin programmer is used in conjunction with specific programmer software to
interrogate and program the implanted device at implant and during patient follow-up
procedures.
Remote Monitoring:
PLATINIUM SonR CRT-D is also equipped with the RF wireless technology which enables to
remotely monitor the patients who have the Sorin SMARTVIEW Monitor installed at home.
30
SORIN PLATINIUM SonR CRT-D U463A