User's Manual
US-ENGLISH – 85
21.4. REMOTE ALERTS AND WARNINGS
The device routinely performs security self-checks and technical
measurements to ensure system integrity. When system integrity is
found to be at risk outside a follow-up, alerts are stored in the device
memory. When system integrity is found to be at risk during a follow-
up, the information is managed as a warning (pop-up message) to
notify immediately the user. For example, the following types of event
can trigger a warning or an alert: technical problem during a shock,
pacing lead impedance or coil impedance measurements out-of-
range, battery depletion, etc. The Remote tab presents an overview
of all the alerts managed by the device.
General parameters
Values
RF communication (1)
ON-OFF
Remote alerts (1)
ON-OFF
(1) RF and Remote alerts are turned on automatically if Shocks are
programmed ON.
System Alerts
Values
Battery depletion – ERI
ON-OFF
Device reset
ON-OFF
Excessive charge time (>25s)
ON-OFF
System integrity
ON-OFF