User's Manual
74 – US-ENGLISH
Programmed ampl. (V)
3.25*
3.5
3.75*
4
4.25*
4.5
Mid-pulse delivered ampl. (V)
3.05
3.25
3.39
3.58
3.88
4.23
Stored amplitude (V)
3.58
3.82
3.98
4.20
4.55
4.96
Programmed ampl. (V)
4.75*
5
6
7*
Mid-pulse delivered ampl. (V)
4.36
4.47
5.37
6.26
Stored amplitude (V)
5.12
5.25
6.30
7.35
* For left ventricular amplitude only.
Ventricular arrhythmia
prevention
Values
Atrial pacing on PVC
Yes-No
Post extrasystolic pause suppression
Yes-No
Acceleration on PVC
Yes-No
Max accelerated rate (min-1)
From 60 to 145 by steps of 5; 100