User's Manual

US-ENGLISH 73
Pacing/Sensing
Values
Atrial sensitivity (mV)
(1)
From 0.2 to 4 by steps of 0.2; 0.4 50 %)
Atrial amplitude (V)
(2)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
Atrial pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
Ventricular sensitivity (mV)
(1)
From 0.4 to 4 by steps of 0.2; 0.4 50 %)
RV amplitude (V)
(2)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 20 %)
RV pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
LV amplitude (V)
(2)
0.25 (± 50 %)
0.5-0.75- 30 %)
1-1.25-1.5-1.75-2-2.25-2.5-2.75-3-3.25-3.5-
3.75-4-4.25-4.5-4.75-5-6-7 (± 20 %)
LV pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
LV pacing polarity
LV bipolar-LV tip to RV-LV ring to RV coil
V chambers
Right-R+L-L+R
VV delay (ms)
0-8-16-24-32-40-48-56-64 (± 3 ms)
(1) Values are measured using a positive and negative triangular signal of
2/13 ms.
(2) The correlation between the programmed amplitudes, the stored
amplitudes and the mid-pulse delivered amplitudes under a 500 ohm load are
given in the following table:
Programmed ampl. (V)
0.25*
0.5*
0.75*
1
1.25*
1.5
Mid-pulse delivered ampl. (V)
0.28
0.49
0.76
0.97
1.18
1.39
Stored amplitude (V)
0.33
0.57
0.89
1.14
1.38
1.63
Programmed ampl. (V)
1.75*
2
2.25*
2.5
2.75*
3
Mid-pulse delivered ampl. (V)
1.66
1.79
2.08
2.35
2.56
2.84
Stored amplitude (V)
1.95
2.10
2.44
2.76
3.01
3.33