User's Manual
US-ENGLISH – 69
20.2. BATTERY
Manufacturer
Greatbatch
Type
Quasar High Rate (QHR)
Model
GB 2593
Number of batteries
1
Total capacity
1964 mAh
Usable capacity
Between BOL and ERI: 1278 mAh.
Between BOL and EOL: 1675 mAh.
Voltage
BOL: 3.25 V. ERI: 2.66 V. EOL: 2.5 V.
20.3. LONGEVITY
The longevities mentioned below are calculated by taking into
account 6 months storage.
5.1 years
Biventricular pacing in DDD mode, 100%, 500 ohm, 3.5 V, 0.35 ms,
60 min-1, one 42 J shock per quarter, sensors OFF
4.6 years
Biventricular pacing in DDD mode, 100%, 500 ohm, 3.5 V, 0.35 ms,
60 min-1, one 42 J shock per quarter, sensors (G, SonR) ON
6.0 years
Biventricular pacing in DDD mode, 1% in atrium, 100% in both
ventricles, 500 ohm, 3.5 V, 0.35 ms, 60 min-1, one 42 J shock per
quarter, sensors OFF
4.2 years
Biventricular pacing in DDD mode, 15% in atrium, 100% in both
ventricles, 500 ohm, 4.5 V, 0.50 ms, 60 min-1, one 42 J shock per
quarter, sensors OFF
9.0 years
0% pacing, one 42 J shock per quarter, sensors OFF