User's Manual

US-ENGLISH 55
17.2. HOLTER FUNCTION
The Holter records up to 14 tachyarrhythmia episodes as well as the
therapy history.
STORED EPISODES
PARADYM RF SonR CRT-D 9770 stores up to 14 episodes (VF, VT,
Slow VT, SVT/ST, non-sustained).
For each episode four levels of details are presented:
Tachogram (to visualize PP and PR intervals)
Event log for the entire episode:
PARAD/PARAD+ analysis for each majority,
Delivered therapies,
Markers: Atrial, ventricular and biventricular markers, sensed,
paced and in relative refractory periods,
EGM: onset and detection of the arrhythmia, on two therapies,
and the return to slow rhythm by recording electrogram.
Therapy history: For each arrhythmia detection, each therapy
delivered (either automatically or during an electrophysiological
study) and at the end of each arrhythmia, PARADYM RF SonR CRT-
D 9770 records the type of majority rhythm, the number of ATP
sequences delivered, the energy and the number of shocks
delivered.