User's Manual
US-ENGLISH – 53
Alert system
The following set of alert trigger can be independently programmed
ON/OFF by the physician using the Orchestra Plus programmer and
can trigger an alert transmission:
─ Reset of the device
─ ERI reached
─ Low or high impedance (A, RV, LV)
─ Abnormal coil impedance (shock lead)
─ Low or High shock impedance
─ Long charge time
─ Inefficient high energy shock
─ All shocks programmed OFF
─ Shock treated VT/VF
─ Lack of V pacing in CRT device
─ Suspicion of noise on the V lead
─ AT/AF occurrence
─ Fast V rate during AT/AF
WARNINGS
The use of remote monitoring does not replace regular follow-up.
Therefore, when using remote monitoring, the time period between
follow-ups visits may not be extended.
When ERI mode is reached, this information is transmitted via the
remote monitoring facility and then the remote-monitoring is switched
off to preserve battery life.