Alert system
The following set of alert trigger can be independently programmed
ON/OFF by the physician using the Orchestra Plus programmer and
can trigger an alert transmission:
Reset of the device
ERI reached
Low or high impedance (A, RV, LV)
Abnormal coil impedance (shock lead)
Low or High shock impedance
Long charge time
Inefficient high energy shock
All shocks programmed OFF
Shock treated VT/VF
Lack of V pacing in CRT device
Suspicion of noise on the V lead
AT/AF occurrence
Fast V rate during AT/AF
WARNINGS
The use of remote monitoring does not replace regular follow-up.
Therefore, when using remote monitoring, the time period between
follow-ups visits may not be extended.
When ERI mode is reached, this information is transmitted via the
remote monitoring facility and then the remote-monitoring is switched
off to preserve battery life.