User's Manual
52 – US-ENGLISH
Data transmitted
The data transmitted are identical to the data available during a
standard interrogation with the Orchestra Plus programmer. All
counters, histograms, IEGMs and diagnosis available in the device
are transmitted containing (not exhaustive list):
─ programmed parameters
─ Information on patient and system implanted
─ battery status
─ lead status (brady leads and defibrillation coils)
─ pacing counters and mean heart rate (brady)
─ atrial and ventricular arrhythmia counters and episodes
─ ventricular therapy counters
─ heart failure monitoring
Data are presented in the form of 2 reports to the physician: the first
one contains a summary of major counters, histograms, warnings
and diagnosis. The second one presents the 3 most important IEGM
episodes automatically selected based on the degree of severity for
the patient.
User website
On the website, the physician is able to:
─ consult and schedule the remote follow-ups of their patient
─ configure additional ways of being notified of alerts (for instance by
SMS, fax or e-mail
─ consult, print and export patient reports