User's Manual

US-ENGLISH 49
16.7. FOLLOW-UP FUNCTION
Storage of memory data: AIDA+ (Automatic Interpretation for
Diagnosis Assistance) software provides access up to 6 months of
patient follow-up with day by day data collection, or up to 24 hours
with hourly data collection. Episodes of ventricular tachyarrhythmia
are recorded with the programmable EGM channels: either by
selecting up to two traces, or by selecting "Double V" which enables
a one-channel recording that is twice as long.
Alerts / Warnings: The device routinely performs security self-
checks and technical measurements to ensure system integrity.
When system integrity is found to be at risk outside a follow-up, alerts
are stored in the device memory. When system integrity is found to
be at risk during a follow-up, the information is managed as a
warning (pop-up message) to notify immediately the user. For
example, the following types of event can trigger a warning or an
alert: technical problem during a shock, pacing lead impedance or
coil impedance measurements out-of-range, battery depletion,…