14.11. TESTS AND PROGRAMMING
During the implant testing procedure, it is recommended that a
security margin of at least 10 J be demonstrated between the
effective shock energy and maximum programmable energy.
Enable shock therapies, then program the defibrillator.
Verify that the defibrillation lead impedance for each shock delivered
ranges from 30 to 150 ohms. Check the lead connection if the values are
outside these boundaries.
Save the programming data on the programmer’s hard disk and on an
external storage device (if desired).
15. SPECIAL MODES
15.1. SAFETY MODE (NOMINAL VALUES)
Nominal values may be rapidly restored by pressing the following
button on the programming head or programmer keyboard:
or via the “Emergency” button on the SMARTVIEW screen.
In safety mode, the defibrillator operates with the parameters
underlined in the table of programmable parameters.
15.2. MAGNET MODE
When the magnet is applied:
─ antiarrhythmia functions are inhibited (detection of rhythm
disturbances, charging, and therapy),
─ hysteresis, VV delay and AVD paced/sensed offset are set to 0,
─ pacing amplitude is set to 6 V,
─ pulse width is set to maximum,