User's Manual
US-ENGLISH – 39
Connectors: The ventricular connectors are compatible with the IS-1
standard and the defibrillation connectors are compatible with the
DF-1 standard (refer to the “Lead evaluation and lead connection”
sub-section in the “Warnings and precautions” section). Shock
configuration (+ -> -)
The shock configuration is the energy pathway between the
defibrillation electrodes. If an atrial coil is present, the shock
configuration can be programmed for bi-directional shocks.
Programming: When active case and SVC are both programmed to
Yes, the shock configuration can be programmed to:
─ RV to Case (or Case to RV),
─ or RV to SVC (or SVC to RV),
─ or RV to Case+SVC (or Case+SVC to RV).
The polarity of shock is determined by the parameter itself.
14.8. MEASUREMENT OF THRESHOLDS AT
IMPLANT
Pacing and sensing thresholds should be measured at implant.
RV to Case+SVC
RV to Case
RV to SVC