User's Manual
US-ENGLISH – 35
─ Data storage:
─ Therapy History Report,
─ Statistics (pace/sense, therapy, shocks, and battery voltage),
─ Up to 14 complete Holter records with event logs, marker
channel notation, and electrogram records.
The connector head has five ports:
─ Atrial “sonR” port: performs atrial bipolar pace/sense if a
conventional IS-1 lead is connected.
─ RV “IS-1” port: performs right ventricular bipolar pace/sense.
─ LV “IS-1” port: performs left ventricular bipolar pace.
─ RV “DF-1” ports for RV defibrillation coil.
─ SVC “DF-1” port for SVC defibrillation coil.
Distal lead terminal connections are secured with set-screws accessed
via self-sealing silicone plugs. All lead connections pass through the
header into the device via feedthroughs.
Programming System: The Sorin CRM programmer is used in
conjunction with specific programmer software to interrogate and
program the implanted device at implant and during patient follow-up
procedures.
14. IMPLANT PROCEDURE
14.1. NECESSARY EQUIPMENT
Implantation of PARADYM RF SonR CRT-D 9770 requires the
following equipment:
─ Sorin CRM Dedicated programmer, equipped with the
SMARTVIEW software interface and with the programming head,
─ pacing system analyzer, as well as its sterile connecting cables,
to evaluate the pacing and sensing thresholds,
─ a complete set of leads with corresponding introducers,