Shock program
Values
Shock 1 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-
7-8-9- (± 30 %)
10-12-14-16-18-20-22-24-26-28-30-32-
34-42 (± 15 %)
Shock 2 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-
7-8-9- (± 30 %)
10-12-14-16-18-20-22-24-26-28-30-32-
34-42 (± 15 %)
Number of Max. Shock (42 J)
1-2-3-4
21.4. REMOTE ALERTS AND WARNINGS
General parameters
Values
RF communication
(1)
ON-OFF
Remote alerts
(1)
ON-OFF
(1) RF and Remote alerts are turned on automatically if Shocks are programmed
ON.
System Alerts
Values
Battery depletion – ERI
ON-OFF
Device reset
ON-OFF
Excessive charge time (>25s)
ON-OFF
System integrity
ON-OFF