User's Manual
US-ENGLISH – 71
(1) Values are measured using a positive and negative triangular signal of 2/13
ms.
(2) The correlation between the programmed amplitudes, the stored amplitudes
and the mid-pulse delivered amplitudes under a 500 ohm load are given in
the following table:
Programmed ampl. (V)
0.25*
0.5*
0.75*
1
1.25*
1.5
Mid-pulse delivered ampl. (V)
0.28
0.49
0.76
0.97
1.18
1.39
Stored amplitude (V)
0.33
0.57
0.89
1.14
1.38
1.63
Programmed ampl. (V)
1.75*
2
2.25*
2.5
2.75*
3
Mid-pulse delivered ampl. (V)
1.66
1.79
2.08
2.35
2.56
2.84
Stored amplitude (V)
1.95
2.10
2.44
2.76
3.01
3.33
Programmed ampl. (V)
3.25*
3.5
3.75*
4
4.25*
4.5
Mid-pulse delivered ampl. (V)
3.05
3.25
3.39
3.58
3.88
4.23
Stored amplitude (V)
3.58
3.82
3.98
4.20
4.55
4.96
Programmed ampl. (V)
4.75*
5
6
7*
Mid-pulse delivered ampl. (V)
4.36
4.47
5.37
6.26
Stored amplitude (V)
5.12
5.25
6.30
7.35
* For left ventricular amplitude only.
Ventricular arrhythmia
prevention
Values
Atrial pacing on PVC
Yes-No
Post extrasystolic pause suppression
Yes-No
Acceleration on PVC
Yes-No
Max accelerated rate (min-1)
From 60 to 145 by steps of 5; 100